Inserting NMR Sample: Difference between revisions
No edit summary |
No edit summary |
||
Line 3: | Line 3: | ||
== Adjusting sample depth with a sample gauge == | == Adjusting sample depth with a sample gauge == | ||
[[Image:Gauge sch.jpg|right|100px|Shigemi tube]] | [[Image:Gauge sch.jpg|right|100px|Shigemi tube]]First, the NMR tube has to be inserted into a spinner turbine. For best results, NMR tubes should be positioned in the spinner turbine so that the sample volume will be centered in the NMR probe. This can be done with the help of a sample gauge, which usually has a center line and rectangular markings to indicate the probe geometry.<br> | ||
A sample gauge also has an adjustable bottom, which can be adjusted to allow quick posistioning of a series of identical samples. Setting a sample tube too low may cause it to break, as well as contaminate and damage the probe! | |||
It is also important to make sure that the tube sits firmly in the spinner. If your sample slides too easily inside a spinner turbine, it may indicate that its rubber o-rings need replacement. | |||
It is also important to make sure that the tube sits firmly in the spinner. If your sample slides too easily inside a | |||
== Inserting and ejecting == | == Inserting and ejecting == |
Revision as of 22:23, 3 November 2009
Adjusting sample depth with a sample gauge
First, the NMR tube has to be inserted into a spinner turbine. For best results, NMR tubes should be positioned in the spinner turbine so that the sample volume will be centered in the NMR probe. This can be done with the help of a sample gauge, which usually has a center line and rectangular markings to indicate the probe geometry.
A sample gauge also has an adjustable bottom, which can be adjusted to allow quick posistioning of a series of identical samples. Setting a sample tube too low may cause it to break, as well as contaminate and damage the probe!
It is also important to make sure that the tube sits firmly in the spinner. If your sample slides too easily inside a spinner turbine, it may indicate that its rubber o-rings need replacement.
Inserting and ejecting
In VNMR:
- Click on the Acqi button or type acqi
- Use the Insert and Eject buttons in the Acqi window
In VNMRJ:
- Click on the Setup tab
- Uset the Insert sample and Eject sample buttons in the middle of the screen
You can also type insert or i and eject or e at the VNMR or VNMRJ prompt.
Procedure
- Eject the current sample first. Even if there is no sample in the magnet you need to use eject to create the airflow.
- Remove the ejected sample.
- Verify that you have the airflow to support your sample. Carefully insert the sample in the bore.
- Select insert to lower your sample
Figure2. Acquisition window.
Do not leave the spectrometer with an open bore for long periods of time. If you have to remove a sample and have nothing else to measure, either insert a standard NMR sample or cover the bore with a dedicated plug.
External links
www.varianinc.com/image/vimage/docs/products/nmr/apps/pubs/manuals/0199931300b.pdf
-- Main.GaohuaLiu - 24 Jan 2007