NOE Calibration Using CYANA: Difference between revisions
No edit summary |
No edit summary |
||
(7 intermediate revisions by the same user not shown) | |||
Line 3: | Line 3: | ||
NOE calibration converts NOESY cross-peak intensities into upper distance limits. A certain functional relationship is assumed between peak intensities and UPLs for a given group of peaks. The most common is NOE peak volume proportional to interproton distance to the minus six: <br> | NOE calibration converts NOESY cross-peak intensities into upper distance limits. A certain functional relationship is assumed between peak intensities and UPLs for a given group of peaks. The most common is NOE peak volume proportional to interproton distance to the minus six: <br> | ||
<pre> V = A/d^6</pre> | <pre> V = A/d^6</pre> | ||
where < | where <tt>V</tt> is the peak volume (or intensity), <tt>d</tt> is the upper distance limit, and <tt>A</tt> is the calibration constant. | ||
<br> | <br> | ||
Line 15: | Line 15: | ||
=== '''Command "calibrate"''' === | === '''Command "calibrate"''' === | ||
A CYANA command < | A CYANA command <tt>calibrate</tt> is the basic calibration tool, unchanged since DYANA. | ||
<br> | <br> | ||
Line 23: | Line 23: | ||
plot=file log minimal (default: none) | plot=file log minimal (default: none) | ||
</pre> | </pre> | ||
< | <tt>calibrate</tt> accepts a calibration function <tt>f(d)</tt> as the required argument. | ||
This command uses "center" averaging. Explicit pseudoatom corrections are added to UPLs: | This command uses "center" averaging. Explicit pseudoatom corrections are added to UPLs: | ||
*Multiplicity correction is applied by dividing the peak volume by the numbers of 1H spins in pseudoatoms assigned to the peak. For instance, the volume of a cross peak between a Leu QQD pseudo atom and a Tyr QD pseudo atom is divided by a factor of 6 x 2 = 12 prior to applying the calibration function. The resulting UPL is the subject to the upper and lower cutoffs. | *Multiplicity correction is applied by dividing the peak volume by the numbers of 1H spins in pseudoatoms assigned to the peak. For instance, the volume of a cross peak between a Leu QQD pseudo atom and a Tyr QD pseudo atom is divided by a factor of 6 x 2 = 12 prior to applying the calibration function. The resulting UPL is the subject to the upper and lower cutoffs. | ||
*Distance correction is applied by adding a distance between the pseudoatom and its constituent spins. It is applied after the application of upper and lower cutoffs. For example, for a Tyr QD pseudoatom this correction is equal to half the distance between the HD1 and HD2 spins. Distance correction can be manually performed with the < | *Distance correction is applied by adding a distance between the pseudoatom and its constituent spins. It is applied after the application of upper and lower cutoffs. For example, for a Tyr QD pseudoatom this correction is equal to half the distance between the HD1 and HD2 spins. Distance correction can be manually performed with the <tt>distance correct</tt> command. | ||
Type < | Type <tt>help calibrate</tt> at CYANA prompt for more information or see the DYANA documentation. | ||
<br> | <br> | ||
Line 36: | Line 36: | ||
=== '''Macro "caliba"''' === | === '''Macro "caliba"''' === | ||
CYANA macro < | CYANA macro <tt>caliba.cya</tt> is functionally the same as <tt>caliba.dya</tt> of DYANA. It is a more sophisticated calibration tool aimed at getting more nuanced calibration than a uniform <code>1/d</code><sup><code>6</code></sup> dependence. CALIBA is recommended for manual structure calculation with CYANA 2.1. The <tt>caliba</tt> macro in fact uses <tt>calibrate</tt>, therefore, it also uses "center" averaging. | ||
<br> | <br> | ||
Line 54: | Line 54: | ||
<br> | <br> | ||
{| rules="rows" cellspacing="0" cellpadding="0" border="1" style="width: 638px; height: 76px; | {| rules="rows" cellspacing="0" cellpadding="0" border="1" class="twikiTable" id="table1" style="width: 638px; height: 76px;" | ||
|- class="twikiTableOdd twikiTableRowdataBgSorted0 twikiTableRowdataBg0" | |- class="twikiTableOdd twikiTableRowdataBgSorted0 twikiTableRowdataBg0" | ||
! valign="top" bgcolor="#687684" class="twikiTableCol0 twikiFirstCol" | <font color="#ffffff">class</font> | ! valign="top" bgcolor="#687684" class="twikiTableCol0 twikiFirstCol" | <font color="#ffffff">class</font> | ||
Line 93: | Line 93: | ||
<br> | <br> | ||
By default, constant < | By default, constant <tt>A</tt> is calculated from <tt>avedis</tt> - the assumed average distance between backbone atoms. It is also possible to specify the backbone constant <tt>A</tt> explicitly. Constant <tt>B</tt> is then calculated from <tt>A</tt>, and <tt>C</tt> from <tt>B</tt> | ||
Parameters < | Parameters <tt>dmin</tt> and <tt>dmax</tt> are the lower and upper cutoff limits for UPLs, respectively. | ||
Parameter < | Parameter <tt>vmin</tt> is useful to filter out peaks with intensities below a certain threshold. Without it even incorrectly picked peaks with intensities below S/N level will produce UPLs set at the upper cutoff. | ||
< | <tt>plot</tt> is an optional parameter, used to create logarithmic plots of volumes versus corresponding minimal distances in the selected structures together with the calibration functions. It is useful for fine-tuning the calibration during structure refinement. Of course, a structure must be then loaded into CYANA before running <tt>caliba</tt>. | ||
<br> '''VERY IMPORTANT!!!'''. The default behavior of structure calculation (< | <br> '''VERY IMPORTANT!!!'''. The default behavior of structure calculation (<tt>anneal</tt>, <tt>calc_all</tt>, <tt>noeassign</tt>) in CYANA 2.x is to use "sum of r<sup>-6</sup>" averaging for UPLs. Setting <tt>expand=1</tt> declares "sum of r<sup>-6</sup>" averaging, and <tt>expand=0</tt> declares "center" averaging. If your UPLs reflect "center" averaging (that is, they were created with <tt>caliba</tt> or <tt>calibrate</tt>), then | ||
*They can be used with DYANA, CYANA 1.x and AutoStructure. | *They can be used with DYANA, CYANA 1.x and AutoStructure. | ||
*They can be used with legacy macros in CYANA 2.1, such as < | *They can be used with legacy macros in CYANA 2.1, such as <tt>habas</tt>. | ||
*They can be used for manual structure calculation in CYANA 2.1, if < | *They can be used for manual structure calculation in CYANA 2.1, if <tt>expand=0</tt> is set before calling <code>anneal</code>. | ||
*'''Do not''' use them for automated structure calculation in CYANA 2.1 with < | *'''Do not''' use them for automated structure calculation in CYANA 2.1 with <tt>noeassign</tt>. The <tt>noeassign</tt> macro employs <tt>peaks calibrate</tt> (see below) and sets <tt>expand=1</tt>. In general, '''do not''' mix them with UPLs calibrated with <tt>peaks calibrate</tt>. | ||
Type < | Type <tt>help caliba</tt> at CYANA prompt or see the DYANA manual for more information. You can also look at the script itself in the <tt>macro</tt> subdirectory of a CYANA installation. | ||
<br> | <br> | ||
Line 114: | Line 114: | ||
=== '''Command "peaks calibrate"''' === | === '''Command "peaks calibrate"''' === | ||
Command < | Command <tt>peaks calibrate</tt> is a new feature of CYANA 2.x, which produced a "sum of r<sup>-6</sup>"-averaged calibration. The documentation for it is scarce. For example, the manual structure calculation script <tt>~/demo/manual/CALC.cya</tt> contains these lines: | ||
<br> | <br> | ||
Line 130: | Line 130: | ||
write upl $name-in.upl | write upl $name-in.upl | ||
</pre> | </pre> | ||
Line < | Line <tt>calibration prot=$prot peaks=$peaks constant=$calibration dref=$dref</tt> calls the macro <tt>calibration.cya</tt>, which iteratively sets calibration constants for each peaklist. If the <tt>calibration</tt> list variable is defined, its values are taken as calibration constants. If it is left blank, calibration constants are calibrated based on assumed average distance <tt>dref</tt> between '''all''' atoms of the molecule. Note that <tt>dref</tt> here is inherently different from <tt>avedis</tt> of <tt>caliba</tt> - the latter refers only to distances between <tt>backbone</tt> protons. | ||
Line < | Line <tt>peaks calibrate "**" simple</tt> generates UPLs assuming simple <tt>C/d^6</tt> functions with calibration constants <tt>C</tt> from the previous step. These calibration constants are thus closely related the "backbone" class constants <tt>A</tt> of <tt>caliba</tt>. | ||
To modify upper and lower cutoffs for NOE calibration with < | To modify upper and lower cutoffs for NOE calibration with <tt>peaks calibrate</tt> add a line <tt>set upl_values:=2.4,6.0</tt>. The defaults are 2.4 and 5.5. | ||
The main difference from < | The main difference from <tt>calibrate</tt> and <tt>caliba</tt> is that pseudoatom correction are '''not applied''' to resulting UPLs. | ||
<br> | <br> | ||
'''VERY IMPORTANT!!!'''. The default behavior of structure calculation (< | '''VERY IMPORTANT!!!'''. The default behavior of structure calculation (<tt>anneal</tt>, <tt>calc_all</tt>, <tt>noeassign</tt>) in CYANA 2.x is to use "sum of r<sup>-6</sup>" averaging for UPLs. Setting <tt>expand=1</tt> declares "sum of r<sup>-6</sup>" averaging, and <tt>expand=0</tt> declares "center" averaging. If you UPLs use "sum of r<sup>-6</sup>" averaging (if created with <tt>peaks calibrate</tt>), then | ||
*'''Do not''' use them with DYANA, CYANA 1.x and AutoStructure. | *'''Do not''' use them with DYANA, CYANA 1.x and AutoStructure. | ||
*'''Do not''' use them with legacy macros in CYANA 2.1, such as < | *'''Do not''' use them with legacy macros in CYANA 2.1, such as <tt>habas</tt>. | ||
*They can be used for manual structure calculation in CYANA 2.1, if < | *They can be used for manual structure calculation in CYANA 2.1, if <tt>expand=1</tt> is set before calling <tt>anneal</tt>. | ||
*They can be used for automated structure calculation in CYANA 2.1 with < | *They can be used for automated structure calculation in CYANA 2.1 with <tt>noeassign</tt>. The <tt>noeassign</tt> macro employs <tt>peaks calibrate</tt> (see below) and sets <tt>expand=1</tt>. | ||
<br> | <br> | ||
The aim of this new calibration method is likely to facilitate automated NOE assignment and structure calculation with < | The aim of this new calibration method is likely to facilitate automated NOE assignment and structure calculation with <tt>noeassign</tt>. Unlike <tt>caliba</tt> and <tt>calibrate</tt>, <tt>peaks calibrate</tt> effectively generates UPLs even for unassigned peaks assuming NOEs between single protons. If such a peak gets assigned to a pseudoatom, the required correction is applied implicitly during the calculation. | ||
<br> | <br> | ||
Line 155: | Line 155: | ||
=== Initial NOE Calibration === | === Initial NOE Calibration === | ||
After simulating and filtering one typically has three initial NOESY peaklists, for example, < | After simulating and filtering one typically has three initial NOESY peaklists, for example, <tt>n.peaks</tt>, <tt>ali.peaks</tt>, and <tt>aro.peaks</tt>. The assigned peaks are intra-residue and short-range within secondary structure elements. Initial calibration is needed to verify these assignments for consistency with molecular geometry and, if possible, to derive stereospecific assignments with HABAS in CYANA. If your SequenceList and AtomList were generated with XEASY, you will need to convert them to the format of CYANA 2.1 (see the <tt>~/demo/details/MigrateFromDyanaCyana1.cya</tt> example script). SequenceList and AtomList from CARA are already compatible. | ||
Amide, aliphatic and aromatic protons tend to have different relaxation properties in a simultaneous 3D | Amide, aliphatic and aromatic protons tend to have different relaxation properties in a simultaneous 3D <sup>15</sup>N-, <sup>13</sup>C-resolved NOESY. NOESY spectra acquired separately may have varying acquisition times. Therefore, it is recommended to run <tt>caliba</tt> separately for each peaklist. | ||
Here is a sample [[Media:Run_caliba.cya|run_caliba.cya]] script for running < | Here is a sample [[Media:Run_caliba.cya|run_caliba.cya]] script for running <tt>caliba</tt> on all three peaklists: | ||
<pre>protocol:=caliba.log | <pre>protocol:=caliba.log | ||
read prot $name | read prot $name | ||
Line 180: | Line 180: | ||
write upl short.upl | write upl short.upl | ||
protocol:=</pre> | protocol:=</pre> | ||
Note that you need to convert volumes to absolute values before calling < | Note that you need to convert volumes to absolute values before calling <tt>caliba</tt>, otherwise negative intensities will be misinterpreted. | ||
Macro < | Macro <tt>caliba</tt> should work fine for <tt>n.peaks</tt> and <tt>ali.peaks</tt> with default parameters. Aromatic peaklist <tt>aro.peaks</tt> requires special treatment because it does not have any peaks of the backbone class. '''For aromatic peaklists <tt>caliba</tt> should be run either with a single <tt>bb</tt> parameter or with both <tt>bb</tt> and <tt>sc</tt> parameters. Calling it without <tt>bb</tt> parameter will cause an error.''' | ||
The < | The <tt>sc</tt> and <tt>bb</tt> parameters for <tt>aro.peaks</tt> can be estimated from the aliphatic calibration. Typically the aromatic part of the simultaneous NOESY spectrum is less sensitive than the aromatic, therefore the values should be somewhat lower. | ||
<br> | <br> | ||
Line 196: | Line 196: | ||
<br> | <br> | ||
{| rules="rows" cellspacing="0" cellpadding="0" border="1" style="width: 556px; height: 151px; | {| rules="rows" cellspacing="0" cellpadding="0" border="1" class="twikiTable" id="table2" style="width: 556px; height: 151px;" | ||
|- class="twikiTableOdd twikiTableRowdataBgSorted0 twikiTableRowdataBg0" | |- class="twikiTableOdd twikiTableRowdataBgSorted0 twikiTableRowdataBg0" | ||
! valign="top" bgcolor="#687684" class="twikiTableCol0 twikiFirstCol" | | ! valign="top" bgcolor="#687684" class="twikiTableCol0 twikiFirstCol" | | ||
Line 299: | Line 299: | ||
<br> | <br> | ||
{| rules="rows" cellspacing="0" cellpadding="0" border="1" style="width: 285px; height: 106px; | {| rules="rows" cellspacing="0" cellpadding="0" border="1" class="twikiTable" id="table3" style="width: 285px; height: 106px;" | ||
|- class="twikiTableOdd twikiTableRowdataBgSorted0 twikiTableRowdataBg0" | |- class="twikiTableOdd twikiTableRowdataBgSorted0 twikiTableRowdataBg0" | ||
| valign="top" bgcolor="#ffffff" class="twikiTableCol0 twikiFirstCol" | <br> | | valign="top" bgcolor="#ffffff" class="twikiTableCol0 twikiFirstCol" | <br> | ||
Line 334: | Line 334: | ||
<br> | <br> | ||
<br> | |||
== '''References''' == | |||
1. Wuthrich, K. (1986) "NMR of Proteins and Nucleic Acids", page 127. | |||
<br> | |||
<br> |
Latest revision as of 21:53, 6 January 2010
Introduction
NOE calibration converts NOESY cross-peak intensities into upper distance limits. A certain functional relationship is assumed between peak intensities and UPLs for a given group of peaks. The most common is NOE peak volume proportional to interproton distance to the minus six:
V = A/d^6
where V is the peak volume (or intensity), d is the upper distance limit, and A is the calibration constant.
NOE Calibration Using CYANA 2.1
This section describes how to determine NOE calibration constants using CYANA 2.1.
Before running the calibration, the N chemical shifts in the peak list must be updated, and the peak lists separated into a list of N noes and C noes, using the UBNMR macro, macro1.
Command "calibrate"
A CYANA command calibrate is the basic calibration tool, unchanged since DYANA.
Parameters: f(d) (required) dmin [d2 ...] dmax (default: 2.4 5.5) weight=w (default: 1.0) plot=file log minimal (default: none)
calibrate accepts a calibration function f(d) as the required argument.
This command uses "center" averaging. Explicit pseudoatom corrections are added to UPLs:
- Multiplicity correction is applied by dividing the peak volume by the numbers of 1H spins in pseudoatoms assigned to the peak. For instance, the volume of a cross peak between a Leu QQD pseudo atom and a Tyr QD pseudo atom is divided by a factor of 6 x 2 = 12 prior to applying the calibration function. The resulting UPL is the subject to the upper and lower cutoffs.
- Distance correction is applied by adding a distance between the pseudoatom and its constituent spins. It is applied after the application of upper and lower cutoffs. For example, for a Tyr QD pseudoatom this correction is equal to half the distance between the HD1 and HD2 spins. Distance correction can be manually performed with the distance correct command.
Type help calibrate at CYANA prompt for more information or see the DYANA documentation.
Macro "caliba"
CYANA macro caliba.cya is functionally the same as caliba.dya of DYANA. It is a more sophisticated calibration tool aimed at getting more nuanced calibration than a uniform 1/d
6
dependence. CALIBA is recommended for manual structure calculation with CYANA 2.1. The caliba macro in fact uses calibrate, therefore, it also uses "center" averaging.
Caliba parameters: dmin=dmin (default: 2.4) dmax=dmax (default: 5.5) vmin=Vmin (default: 0.0) bb=A (default: calculated automatically) sc=B (default: A/dmin2) methyl=C (default: B/3) weight=w (default: 1.0) avedis=d (default: 3.4) plot=file (optional)
This macro divides each peaklist into three classes of peaks with different calibration functions and constants:
class | peaks/constraints | function |
---|---|---|
backbone |
all HN/HA— HN/HA, and HN(i)/HA(i) — HB(j) with abs(i - j) < 5 |
|
sidechain |
neither "backbone" nor "methyl" |
|
methyl |
all involving methyl groups |
|
By default, constant A is calculated from avedis - the assumed average distance between backbone atoms. It is also possible to specify the backbone constant A explicitly. Constant B is then calculated from A, and C from B
Parameters dmin and dmax are the lower and upper cutoff limits for UPLs, respectively.
Parameter vmin is useful to filter out peaks with intensities below a certain threshold. Without it even incorrectly picked peaks with intensities below S/N level will produce UPLs set at the upper cutoff.
plot is an optional parameter, used to create logarithmic plots of volumes versus corresponding minimal distances in the selected structures together with the calibration functions. It is useful for fine-tuning the calibration during structure refinement. Of course, a structure must be then loaded into CYANA before running caliba.
VERY IMPORTANT!!!. The default behavior of structure calculation (anneal, calc_all, noeassign) in CYANA 2.x is to use "sum of r-6" averaging for UPLs. Setting expand=1 declares "sum of r-6" averaging, and expand=0 declares "center" averaging. If your UPLs reflect "center" averaging (that is, they were created with caliba or calibrate), then
- They can be used with DYANA, CYANA 1.x and AutoStructure.
- They can be used with legacy macros in CYANA 2.1, such as habas.
- They can be used for manual structure calculation in CYANA 2.1, if expand=0 is set before calling
anneal
. - Do not use them for automated structure calculation in CYANA 2.1 with noeassign. The noeassign macro employs peaks calibrate (see below) and sets expand=1. In general, do not mix them with UPLs calibrated with peaks calibrate.
Type help caliba at CYANA prompt or see the DYANA manual for more information. You can also look at the script itself in the macro subdirectory of a CYANA installation.
Command "peaks calibrate"
Command peaks calibrate is a new feature of CYANA 2.x, which produced a "sum of r-6"-averaged calibration. The documentation for it is scarce. For example, the manual structure calculation script ~/demo/manual/CALC.cya contains these lines:
peaks := c13,n15,aro # names of peak lists prot := demo # names of proton lists tolerance := 0.040,0.030,0.45 # chemical shift tolerances # order: 1H(a), 1H(b), 13C/15N(b), 13C/15N(a) calibration:= 6.7E5,8.2E5,8.0E4 # calibration constants (will be determined # automatically, if commented out) dref := 4.2 # average upper distance limit for # automatic calibration calibration prot=$prot peaks=$peaks constant=$calibration dref=$dref peaks calibrate "**" simple write upl $name-in.upl
Line calibration prot=$prot peaks=$peaks constant=$calibration dref=$dref calls the macro calibration.cya, which iteratively sets calibration constants for each peaklist. If the calibration list variable is defined, its values are taken as calibration constants. If it is left blank, calibration constants are calibrated based on assumed average distance dref between all atoms of the molecule. Note that dref here is inherently different from avedis of caliba - the latter refers only to distances between backbone protons.
Line peaks calibrate "**" simple generates UPLs assuming simple C/d^6 functions with calibration constants C from the previous step. These calibration constants are thus closely related the "backbone" class constants A of caliba.
To modify upper and lower cutoffs for NOE calibration with peaks calibrate add a line set upl_values:=2.4,6.0. The defaults are 2.4 and 5.5.
The main difference from calibrate and caliba is that pseudoatom correction are not applied to resulting UPLs.
VERY IMPORTANT!!!. The default behavior of structure calculation (anneal, calc_all, noeassign) in CYANA 2.x is to use "sum of r-6" averaging for UPLs. Setting expand=1 declares "sum of r-6" averaging, and expand=0 declares "center" averaging. If you UPLs use "sum of r-6" averaging (if created with peaks calibrate), then
- Do not use them with DYANA, CYANA 1.x and AutoStructure.
- Do not use them with legacy macros in CYANA 2.1, such as habas.
- They can be used for manual structure calculation in CYANA 2.1, if expand=1 is set before calling anneal.
- They can be used for automated structure calculation in CYANA 2.1 with noeassign. The noeassign macro employs peaks calibrate (see below) and sets expand=1.
The aim of this new calibration method is likely to facilitate automated NOE assignment and structure calculation with noeassign. Unlike caliba and calibrate, peaks calibrate effectively generates UPLs even for unassigned peaks assuming NOEs between single protons. If such a peak gets assigned to a pseudoatom, the required correction is applied implicitly during the calculation.
Initial NOE Calibration
After simulating and filtering one typically has three initial NOESY peaklists, for example, n.peaks, ali.peaks, and aro.peaks. The assigned peaks are intra-residue and short-range within secondary structure elements. Initial calibration is needed to verify these assignments for consistency with molecular geometry and, if possible, to derive stereospecific assignments with HABAS in CYANA. If your SequenceList and AtomList were generated with XEASY, you will need to convert them to the format of CYANA 2.1 (see the ~/demo/details/MigrateFromDyanaCyana1.cya example script). SequenceList and AtomList from CARA are already compatible.
Amide, aliphatic and aromatic protons tend to have different relaxation properties in a simultaneous 3D 15N-, 13C-resolved NOESY. NOESY spectra acquired separately may have varying acquisition times. Therefore, it is recommended to run caliba separately for each peaklist.
Here is a sample run_caliba.cya script for running caliba on all three peaklists:
protocol:=caliba.log read prot $name read peaks n peaks select "**" peaks set volume=abs(volume) caliba read peaks ali peaks select "**" peaks set volume=abs(volume) caliba read peaks aro peaks select "**" peaks set volume=abs(volume) caliba bb=1.0E+06 write upl short.upl protocol:=
Note that you need to convert volumes to absolute values before calling caliba, otherwise negative intensities will be misinterpreted.
Macro caliba should work fine for n.peaks and ali.peaks with default parameters. Aromatic peaklist aro.peaks requires special treatment because it does not have any peaks of the backbone class. For aromatic peaklists caliba should be run either with a single bb parameter or with both bb and sc parameters. Calling it without bb parameter will cause an error.
The sc and bb parameters for aro.peaks can be estimated from the aliphatic calibration. Typically the aromatic part of the simultaneous NOESY spectrum is less sensitive than the aromatic, therefore the values should be somewhat lower.
Verifying Initial NOE Calibration
Inspect resulting UPL files to make sure the the calibration is correct. You should avoid making UPLs too tight or too loose.
Known short-range distances in the secondary structure elements constitute good reference points. Consult the table below (adapted in short form from Ref. 1).
distance |
α-helix |
3_10-helix |
β-sheet (A) |
|
---|---|---|---|---|
d_αN |
3.5 |
3.4 |
2.2 |
2.2 |
d_αN(i,i+2) |
4.4 |
3.8 |
||
d_αN(i,i+3) |
3.4 |
3.3 |
||
d_αN(i,i+4) |
4.2 |
3.3 |
||
d_NN |
2.8 |
2.6 |
4.3 |
4.2 |
d_NN(i,i+2) |
4.2 |
4.1 |
For calibration of aromatic peaklists the characteristic distances are:
distance |
center averaging | |
---|---|---|
d_δβ |
2.33 - 3.77 |
5.00 - 6.75 |
d_δε |
2.48 |
7.4 |
Note that aromatic QD and QE are pseudoatoms and the constraints in "center" averaging mode are much larger than the distances between individual atoms.
References
1. Wuthrich, K. (1986) "NMR of Proteins and Nucleic Acids", page 127.