Measuring protein concentration using NanoDrop
Procedure (used at Rutger's University)
Procedure (University of Toronto)
No edit summary |
No edit summary |
||
Line 7: | Line 7: | ||
<br> | <br> | ||
Typical protein | Typical protein concentration measurement protocol with NanoDrop-2000 with connected PC computer. | ||
<br> | <br> | ||
Line 25: | Line 25: | ||
#Clean the sample surfaces once more after you are finished. | #Clean the sample surfaces once more after you are finished. | ||
<br> | <br> [[Image:Nanodrop.jpg]] | ||
[[ | |||
=== Procedure (University of Toronto)<br> === | === Procedure (University of Toronto)<br> === |
Revision as of 22:26, 4 December 2009
Measuring protein concentration using NanoDrop
Procedure (used at Rutger's University)
concentration = Absorbance at 280 nm divided by absorbance coefficient
Typical protein concentration measurement protocol with NanoDrop-2000 with connected PC computer.
- Before starting the software module, clean the sample surfaces with DI water to remove any dried sample that might be present.
- Alternatively, you can clean the sample surfaces with a Kimwipe moistened with 70% ethanol.
- Open the Nanodrop program and the appropriate module (e.g., Protein).
- Wipe off the top and bottom sensors of the instrument with a Kimwipe. These are just the polished ends of fiber optic cable, so wiping is sufficient to prevent carryover.
- Pipette 2 μL of DD water onto the sensor. Bring down the lever arm.
- Follow the onscreen prompts to calibrate.
- Wipe the sensors and pipette on 2 μL of the corresponding blank (Buffer solution your protein is in). Bring down the lever arm.
- Follow the onscreen prompts to blank.
- Wipe the sensors and pipette on 2 μL of your protein sample. Bring down the lever arm.
- Click Measure and record the concentration measured.
- For protein, the peak should be at 280 nm
- To test multiple samples, just wipe the sensor in between measurements with a Kimwipe. Recalibration or re-blanking is not necessary.
- Clean the sample surfaces once more after you are finished.
Procedure (University of Toronto)
Since july of 2009, we use thermofisher nanodrop1000 to measure the protein absorbance at 280 nm.
The molecular weight and theoretical extinction coefficient are calculated using expasy.